System for diagnostics of motor | Verified Russian Supplier

System for diagnostics of motor pathology and

Manufacturer: NEUROSOFT LLC
Price: Request Quote

Bulk pricing available

FOB, CIF & EXW terms available

Description

The specialised biofeedback simulator "Stadis" allows for the assessment of walking parameters and rehabilitation based on the results of the examination. The Stadis simulator uses Neurosens biometric sensors to record walking parameters. These miniature devices are mounted on the patient's shins (and/or back) and record rotational velocities and accelerations in three axes, as well as electromyograms in two channels.

Up to 20 walking parameters can be recorded in total. The sensors are wireless, battery-powered and transmit data to the software via Wi-Fi. They are not affected by nearby metal structures.

System for the diagnosis of walking function. The system for the objective assessment of walking function allows to easily and quickly assess the state of function according to the main groups of parameters (temporal, spatial, kinematic and EMG). Up to 6 inertial sensors can be used simultaneously.

Specifications

Appointment of the device
Designed to analyse the biomechanical/neuromuscular function of walking using specialised equipment
Method of motion registration
Biometric sensors
Automatic calculation of walking time parameters, walking phases, spatial parameters of walking, kinematic parameters of walking
availability
Video recording of evaluation procedures
availability
Automatic generation of evaluation report
availability
Сохранение выполненных оценок, а также их протоколов в БД
availability
Possibility to change the list of calculated parameters for biomechanical/neuromuscular function analysis
availability
Ability to view previous walking function assessments and their protocols.
availability
Ability to create custom templates for walking function assessment protocols
availability
Built-in protocol template editor
availability
Ability to edit evaluation protocols
availability
Ability to include graphs, figures and tables in the protocols
availability
Availability of a consolidated protocol to track the dynamics of rehabilitation
availability
Base of medical norms on basic parameters of walking (phases of walking, temporal parameters, spatial parameters, kinematics)
availability
Monitoring of equipment condition during evaluation procedure
availability
Types of registration channels
accelerometric, gyroscopic, electromyographic
Dimensions of the sensor
Not more than 60×40×17 (± 2) mm
Wireless communication with PC
availability
communication range
Not less than 10 metres in line of sight
Sensor synchronization accuracy
max. 0.005 s discrepancy
Quantization frequency
at least 200 Hz
Input range of measured accelerations in each of the three axes
At least ±16g
Permissible error of acceleration measurement in each of the three axes
not more than ±(0.04 g + 5% of the measured value)
Input range of angular velocities in each of the three axes
Not less than ± 2000 deg/sec.
Permissible error of angular velocity measurement in each of the three axes
not more than ±(0.5 deg/s + 2% of measured value)
Total number of differential channels per 1 sensor
At least 2
Quantization frequency
at least 2 kHz
In-phase noise reduction (CMRR)
At least 100 dB
Weight of sensor, not more
40 year
Lower bandwidth limit at level (-3 ± 0.5) dB in EMG measurement mode
max. 5 Hz
Upper limit of the bandwidth at the level of (-3 ± 0.5) dB in the EMG measurement mode
at least 500 Hz
Natural noise level in EMG measurement mode
Not more than 2 µV RMS
Input differential resistance
At least 100 megohms
EMG signal input range
max. 0.07 µV - 20 mV
EMG signal measurement error in the input range from 0.07 µV to 20 mV and at the bandwidth from 5 to 500 Hz
not more than ±(7 ?V + 5% of the measured value)
Suppression of mains interference by a switchable tap selector filter
At least 40 dB
Function for determining the sub-electrode impedance
availability
Maximum number of sensors simultaneously connected to one PC, not less than
B1/B8 2100/900 MHz pcs

Note: All specifications are provided by the manufacturer and may be subject to change. Please contact us to confirm the latest specifications before placing your order.

Save Your Time

Share your requirements for a quick response!

Instant response in 15 minutes
Best wholesale prices guaranteed
Direct from manufacturer

Delivery & Payment

Shipping Terms

FOB Novorossiysk, Russia CIF Available to major ports worldwide EXW Manufacturer's facility, Russia

Delivery Time

Sea freight: 30-60 days (depending on destination) Air freight: 14-21 days (for urgent orders)

Payment Methods

Letter of Credit (L/C) Wire Transfer (T/T) Escrow Services

Similar Products You May Be Interested In

Verified Suppliers

All products are sourced directly from authorized Russian manufacturers

Quality Assurance

Products meet international quality standards with proper certification

Global Shipping

Reliable logistics solutions to deliver products to your location

Secure Payments

Multiple secure payment options to facilitate international transactions

Your callback request has been received. We'll contact you shortly.